Chauvin Arnoux C.A 1110 Logger Luxmeter

- Measurement ranges : 0.1 lx to 200,000 lx (lux) ; 0.01 fc to 18,580 fc  
- Resolution : 
+ 0.1 lx (0.1 lx to 999.9 lx) / 0.01 fc (0.01 fc to 99.99 fc) 
+ 1 lx (1 000 lx to 9,999 lx) / 0.1 fc (100.0 fc to 999.9 fc) 
+ 10 lx (10,000 lx to 99.99 klx) / 1 fc (1,000 to 9.999 fc) 
+ 100 lx beyond / 10 fc beyond 
METROLOGICAL SPECIFICATIONS
- Intrinsic uncertainty for measurement in lux
+ Class C luxmeter as per the NF C 42-710 standard: ± 3% of reading on incandescent sources (by default) ; ± 6% of reading on LEDs (3,000 K to 6,000 K) ; ± 9% of reading on fluorescent sources
+ Mode with compensation : LED mode: ± 4% of reading (at 4,000 K) ; Fluo mode: ± 4% of reading (type F11, 4,000 K)
FUNCTIONS
- Recording 
+Manual start/stop on the product : Short press for MEM: spot recording / Long press for REC: recording at the rate of the current mode
+ Programmed recording:  Start date, recording interval (from 1 minute to 2 hours) and end date are defined using the PC software
- Data storage: More than 1 million points 
- Min-Avg-Max-Hold : Yes
- MAP mode: The MAP function can be used to map the lighting on a surface or in a room. In this way, the lighting measurements are saved in the same file.
- Units : lx or fc 
- Backlighting: Yes
- Automatic power-off : Yes (configurable) 
POWER SUPPLY
- Type : 3 x 1.5 V AA / LR6 alkaline batteries. Connection to the mains possible with the mains / micro-USB adapter offered as an accessory
- Battery life : 500 hrs (portable mode) / 3 years of recording (15-minute measurement interval) 
PHYSICAL SPECIFICATIONS 
- Interfaces : 2 possible communication modes: Bluetooth wireless link and USB 
- Mounting: Casing equipped with a magnet, a wall-mount system and a slot for suspending the product. Compatible with the Multifix accessory 
- Dimensions / weight :  Casing: 150 x 72 x 32 mm / Sensor: 67 x 64 x 35 mm / 345 g with batteries
- Ingress protection : IP50 
- Product operating range: Temperature: -10 to +60 ° C / Humidity: 0 to 90 %RH 
-Compliance : IEC61010-1 / IEC 61326-1 / Class C luxmeter as per the French NF C 42-710 standard based on the CIE guidelines
GENERAL SPECIFICATIONS 
- Data Logger Transfer Windows PC software functions: Display as a graph or a table of values / Data export in graph or Excel table form. Real-time mode / Automatic report generation in Word format / Data format compatible with the Dataview software offered as an accessory


State at delivery: Bag, 3 x 1.5 V AA alkaline batteries, USB cable, test report and quick start-up guide
Accessories:
- Shockproof sheath + Multifix   P01654252
- Multifix P01102100Z
- Mains adapter  P01651023
- Bag  P01298075
- Metal case  P01298071
- Dataview software P01102095
- BLE Bluetooth modem / USB for PC P01654253 
- 4 x AA/LR6 NiMH Batteries + Battery charger  HX0053

Details


datasheet
user guider

Advantages

  • Spectral error compensation for LED or fluorescent sources
  • MAP function for mapping the illuminance of a surface or room
  • C.I.E. spectral correction and incidence correction
  • Wide backlit display
  • Remote probe limiting the risk of the operator shading the sensor or built-in probe for one-handed use
  • Recording up to 1 million points
  • Communicating via USB or Bluetooth
  • Data Logger Transfer software with automatic report generation

Description

The C.A 1110 luxmeter (light meter) measures the illuminance of all light sources (LED, Fluo. etc.) up to 200,000 lux in compliance with Class C of the NF C 42-710 standard.  
With the MAP function, you can produce a map of a surface or room in order to check that the illuminance is uniform.   
Because it is magnetic, the sensor can be fixed to the product for one-handed use or for easy transport, but it is also possible to use it remotely to avoid shading by the operator.
The C.A 1110 also offers Min, Max, Average and Hold functions, as well as allowing spot or programmable recording.
The Data Logger Transfer software is available for free download to view the recorded data, configure the C.A 1110, benefit from remote display and generate automatic reports. 

 
Additional specifications:

- Measurement range: 0.1 lx to 200,000 lx (lux)
- Intrinsic uncertainty without compensation:
± 3% of the reading on incandescent sources (default)
± 6%  of the reading on LEDs (3,000 K to 6,000 K)
± 9% of the reading on fluorescent sources
- Min, Max, Average, MAP and Hold functions, lx or fc / Backlighting
- Recording up to 1 million points
- USB or Bluetooth interfaces
- IP50 casing
- Casing dimensions: 150 x 72 x 32 mm / Sensor: 67 x 64 x 35 mm
- Weight: 345 g with batteries
- Compatible with the Multifix accessory
- Shockproof protective sheath available as an accessory

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information