Fluke 6200-2 Portable Appliance Tester

Power-on Test
+ Display Range: 90 V to 264 V/ ± (2% + 3 counts)
+ Input Impedance: > 1 MΩ // 2.2 nF
+ Maximum Input Mains Voltage: 264 V
Earth Bond Test (Rpe)
+ Display Range: 0 to 19.99 Ω/ ± (2.5% + 4 counts)
+ Test Current: 200 mA AC -0% +40% into 1.99 Ω; 25 A AC ± 20% into 25 mΩ at 230 V
+ Open Circuit Voltage: > 4 V AC, < 24 V AC
Insulation Test (Riso)
+ Display Range: 0 to 299 MΩ/ ± (5% + 2 counts) from 0.1 to 300 MΩ
+ Test Voltage: 500 V DC -0% + 10% at 500 kΩ load
+ Test Current: > 1 mA at 500 kΩ load, < 15 mA at 0 Ω
+ Auto Discharge Time: < 0.5 s for 1 μF
Touch Current Test
+ Display Range: 0 to 1.99 mA AC/ ± (2.5% + 3 counts)
+ Internal Resistance: 2 kΩ 
Substitute Leakage Current Test
+ Display Range: 0 to 19.99 mA AC/± (2.5% + 3 counts)
+ Test Voltage: 100 V AC ± 20%
Load/Leakage Test
+ Load Current: 0 A to 13 A/± (4% + 2 counts)
PELV Test
+ Accuracy at 50 Hz: ± (2% + 3 counts)
+ Overload Protection:  300 V rms    
+ Warning Threshold: 25 V rms    

Details

Manual

Specifications

Test Specifications

The accuracy specification for the display range is defined as ±(%reading + digit counts) at 23 °C ±5 °C, ≤75 % RH.

Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C.

The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997, EN61557-4: 1997, EN61557-6: 1997, DIN VDE0404-2.

Power-on Test

The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency.

Display Range

90 V to 264 V

Accuracy at 50 Hz

± (2% + 3 counts)

Resolution 

1 V

Input Impedance      

> 1 MΩ // 2.2 nF

Maximum Input Mains Voltage

264 V

Earth Bond Test (Rpe)

Display Range

0 to 19.99 Ω

Accuracy (After Bond Test Zeroing)

± (2.5% + 4 counts)

Resolution

0.01 Ω

Test Current

200 mA AC -0% +40% into 1.99 Ω
25 A AC ± 20% into 25 mΩ at 230 V

Open Circuit Voltage

> 4 V AC, < 24 V AC

Bond Test Zeroing

can subtract up to 1.99 Ω

Insulation Test (Riso)

Display Range

0 to 299 MΩ

Accuracy

± (5% + 2 counts) from 0.1 to 300 MΩ

Resolution

0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 to 199.9 MΩ)
1 MΩ (200 to 299 MΩ)

Test Voltage

500 V DC -0% + 10% at 500 kΩ load

Test Current

> 1 mA at 500 kΩ load, < 15 mA at 0 Ω

Auto Discharge Time

< 0.5 s for 1 μF

Max. Capacitive Load

Operational up to 1 μF

Touch Current Test

Display Range

0 to 1.99 mA AC

Accuracy

± (4% + 2 counts)

Resolution

0.01 mA

Internal Resistance (via Probe)

2 kΩ

Measuring Method

Probe*

Substitute Leakage Current Test

Display Range

0 to 19.99 mA AC

Accuracy

± (2.5% + 3 counts)

Resolution

0.01 mA

Test Voltage

100 V AC ± 20%

Load/Leakage Test: Load Current

Display Range

0 A to 13 A

Accuracy

± (4% + 2 counts)

Resolution

0.1 A*

Load/Leakage Test: Load Power

Display Range 230 V Mains