For full functionality of this site it is necessary to enable JavaScript.

MEDSOURCE TC-600A (auto dry) 蒸気殺菌オートクレーブ (283L)

External  DIA L.W.H. ( cm ): 154 x 76 x 175

Chamber Ø x D ( cm ): 60 x 100

Capacity ( Liter ): 283

PowerConsumption :12KW , 380V-3p

Max SterilizationTemperature ( °C ): 134 ± 2

Operate Pressure ( kg/cm² ): 2.2 ± 0.1

 Option:

Water filling
Safety door lock by pressure
Dry vacuum (TC-750/TC-615)
Dry vacuum (TC-612/TC-600)
Dry vacuum (TC-500/459/409/406/339/336)

Details

Datasheet

Purpose: 

The steam sterilizer is a highly safe sterilizing product with stable function, suitable for major hospitals, clinics and biochemical research laboratories etc., for the sterilization of instrument, cloth, glassware and liquid etc.

Designed of double-layers with inner and outer chamber for good drying efficiency, and totally made of stainless steel 304 # (316 # is optional). The outer layer uses thermal insulation wool for heat preservation, and then covered with 304 # stainless steel sheet for both functions of aesthetic appearance and heat preservation.

Stainless steel door with radial lever can effectively lock tight the door cover and prevent steam leakage.

Silicone door gasket is high- temperature resistant, durable, and without deformation.

The chamber body is supported by a stainless steel support, and the height of level can be adjusted from the adjustable footpad of support base.

Practical Sterilization Function:

  1. Preset Vacuum Function is able to shorten the sterilization time.
  2. Vacuum drying function is able to remove unwanted moisture quickly, and to reach the Bowie Dick test criteria.
  3. Cloth / apparatus glass / liquid, Sterilization Functional Switch Mode Trail Cart is optional can be selected to switch according to different kinds of materials to be sterilized. When liquid sterilization, make sure liquid ingredient to avoid rust and corrosion occurred. For medicalt & chemical ingredient, physiological salt water etc. pls select stainless steel #316
  4. Auto Water Supply Device: Electrical control system is able to detect and sense the water volume automatically. Upon insufficient water volume, it will supply to high water level automatically. Use equipment for soft water purification can prolong the life cycle of sterilizer.

Safety Devices:

  1. Auto Power-Off Protection Device for water shortage, when water runs out, the electrical control system will automatically cut off the heating power to prevent heating device from overheating, dry-out or even meltdown.
  2. Pressure Controller, the maximum pressure in the boiler is limited to 2.2 ± 0.1kg/cm 2. When the pressure reaches the set value, the heating power will be automatically off to prevent the pressure from rising.
  3. Safety Valve, when the pressure reaches 2.5kg/cm 2, it will start a rapid exhaust of steam and release the pressure from the chamber .
  4. Pressure Lock, when the pressure of inner chamber exceeds 0.2kg/cm 2, the clutch will be locked automatically, and the door cover will not be able to open, in order to keep the operator safe.
  5. Safety Switch of Door Cover, when the door cover opened, the steam in the outer chamber is unable to enter the inner chamber. It cannot be operated normally unless the door cover is closed tightly.

Simple Operation Interface:

  1. .Digital LED electronic timer is able to control the working hours of sterilization and drying precisely within the range of 0~99 minutes.
  2. Digital LED electronic temperature controller is able to set the required working temperature quickly, and show the current temperature status of inner chamber. The highest setting temperature is 134°C (218.47 F°)
  3. The pressure gauge of inner and outer chamber is able to clearly show the current working temperature status of interior of sterilizer.
  4. The open and close of turning type door cover is easy for operation.

The highest sterilization temperature is 134°C(218.47 F°), and the pressure value for sterilization is able to reach 2.2kg/cm2 (2.16 bar or 31.29psi).

Heating device with high power will take 20 minutes to complete sterilization under 121°C, or 15 minutes under 132°C.

The preparation steps for sterilization function shown as following:

  • Manual operation- put the items to be sterilized inside, and close the door cover► supply water manually ► set time ► turn on the power ► heating ► sterilization time completed ► exhaust steam manually ► set drying time ► completion
  • Automatic operation- put the items to be sterilized inside, and close the door cover► set temperature/time► turn on the power/activate the switch► supply water manually► heating ► sterilizing► steam exhaust for inner chamber ► drying► completion
  • Fully-automatic operation- put the items to be sterilized inside, and close the door cover► set temperature/time► turn on the power/activate the switch► supply water automatically► heating ►preset vacuum for 3 times ► sterilizing► steam exhaust for inner chamber ► vacuum drying► completion

 

FunctionsOperating system

Manual
drying
function

Automatic
drying
function

Automatic and
vacuum
drying
function

Steam exhaust valve outer chamber

Yes

Yes

Yes

Dry valve, inner chamber

Yes

Yes

Yes

Dry function

Manual

Auto

Auto

Sterilization function

Manual

Auto

Auto

Thermometer, inner chamber

Yes

N/A

N/A

Temperature controller

N/A

Yes

Yes

Timer

60Minutes x 1pcs

90Minutes x 2pcs

99Minutes x 2pcs

Safety valve

1pcs

1pcs

1pcs

Pre-heating function

Yes

Yes

Yes

Auto water filling

Optional

Optional

Yea

Vacuum drying (Pre-vacuum)

N/A

N/A

Yes

Glass & Liquid sterilization function

N/A

Optional

Yea

Low water indication / cut-off

Yes

Yes

Yes

Water Level Control & Protection

Yes

Yes

Yes

Safety door lock activated by pressure

Optional

Optional

Optional

Chamber material stainless steel #316(standard#304)

Optional

Optional

Optional

Trail cart

Optional

Optional

Optional

Stainless steel tray

1pcs

1pcs

1pcs

  • 良質な取り決め
  • オリジナル保証
  • 宅配便
  • 買い取り簡単化

おまけチャンス‐ニュースを受ける登録