KIKUSUI TOS8870A Hipot Tester with Insulation Resistance Test
Hipot Tester
Test Voltage:
Output AC Voltage: 0 V to 2.5 kV/0 V to 5 kV (two ranges)
Output Rating: 500 VA (5 kV, 100 mA with 100 V line voltage) *1
Waveform: AC line waveform
Voltage regulation: Better than 20% (for maximum rated load to no load, with 100 V line voltage)
Switching: With zero-start type switch
Output Voltmeter:
Scales: 2.5 kV f.s / 5 kV f.s, two ranges linear scales
Class of meter: JIS Class 1
Accuracy: 5 °C to 15°C : ±3 % f.s 15 °C to 35 °C : ±1.5 % f.s (with a sine wave ) *2
Indication: Mean-value response, effective-value scale graduation
Test time:
Timer :0.2 s to 99.9 s (× 0.1 range) ±50 ms
1 s to 999 s (× 1 range) ±0.5 s
Others: Terminals for monitoring of leakage current
Insulation Resistance Tester
Measuring Voltage: 500 V or 1000 V DC, negative polarity (two ranges)
Measuring terminal voltage: 0% to + 5% of rated measuring voltage (At rated measuring current or less)
Output current Rated measuring current: 1.0 mA
Short circuit current: 12 mA or less
Effective Measuring:
500 V range: 1 MΩ to 1000 MΩ
Ranges 1000 V range: 2 MΩ to 2000 MΩ
Values center of scale
500 V range: 20 MΩ
1000 V range: 50 MΩ
Accuracy:
1st effective measuring range : ±5 % of the indicated value *1
2nd effective measuring range : ±10 % of the indicated value *1
Dimensions (maximum): 430 (435) W x 132 (155) H x 370 (440) Dmm
Weight Approx. 23 kg
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts