For full functionality of this site it is necessary to enable JavaScript.

Nagman 14+ Temperature Calibrator ( ±0.02% Rdg)

Input Functions:

- DC Voltage (Reference/ Ranges/ Resolution/ Accuracy/Remarks): 

+ 50 mV/ -5 to 55 mV/ 1 mV/ 0.02 + 0.02/Input Resistance 100 MΩ

+ 500 mV/ -50 to 550 mV/ 10 mV/ 0.02 + 0.01/Input Resistance 100 MΩ

- Resistance (Ohm) (Reference/ Ranges/ Resolution/ Accuracy/Remarks): 

+ 500 Ω/ 0 to 550 Ω/ 0.01 Ω/ 0.05 + 0.02/ _

+ 5 KΩ/ 0 KΩ to 5.5 KΩ/0.1 Ω/ 0.05 + 0.02/ _

- Thermocouple (Reference/ Ranges/ Resolution/ Accuracy/Remarks): 

+ R: 0 to 1767°C/ 1°C/ 0 to 500°C: 1.8°C/ _

+ S: 0 to 1767°C/ 1°C/ 500 to 1767°C: 1.5°C/ _

+ K: -100 to 1372°C/ 0.1°C/ -100 to 0°C: 1.2°C; 0 to 1372°C: 0.8°C/ _

+ E: -50 to 850°C/ 0.1°C/ -50 to 0°C: 0.9°C; 0 to 850°C: 1.5°C/ _

+ J: -60 to 1120°C/ 0.1°C/ -60 to 0°C: 1°C; 0 to 1120°C: 0.7°C/ _

+ T: -100 to 400°C/ 0.1°C/ -100 to 0°C: 1°C; 0 to 400°C: 0.7°C/ _

+ N: -200 to 1300°C/ 0.1°C/ -200 to 0°C: 1.5°C; 0 to 1300°C: 0.9°C/ _

+ B: 600 to 1820°C/ 1°C/ 600 to 800°C: 2.2°C; 800 to 1000°C: 1.8°C; 1000 to 1820°C: 1.4°C/ _

+ L: -60 to 900°C/ 0.1°C/ -60 to 0°C: 0.7°C; 0 to 900°C: 0.5°C/ _

+ U: -100 to 600°C/ 0.1°C/ -100 to 0°C: 0.7°C; 0 to 600°C: 0.5°C/ _

-RTD (Reference/ Ranges/ Resolution/ Accuracy):

+ Pt100 385/ -200 to 800°C/ 0.1°C/ -200 to 0°C: 0.5°C

                                 / 0 to 400°C: 0.7°C

                                 /400 to 800°C: 0.8°C

+ Pt1000 385/ -200 to 630°C/ 0.1°C/ -200 to 100°C: 0.8°C

                                 / 100 to 300°C: 0.9°C

                                 /300 to 630°C: 1.0°C

+ Pt200 385/ -200 to 630°C/ 0.1°C/ -200 to 100°C: 0.8°C

                                 / 100 to 300°C: 0.9°C

                                 /300 to 630°C: 1.0°C

+ Pt500 385/ -200 to 630°C/ 0.1°C/ _


+ Cu10/ -100 to 260°C/ 0.1°C/ 1.8°C

+ Cu50/ -50 to 150°C/ 0.1°C/ 0.7°C

Continuity (Reference/ Ranges/ Resolution/ Accuracy/Remarks):

500 Ω/ <50 Ω round/ 0.01 Ω/ _/ Excitation Current approx.. 1 mA


Output Functions: 

- DC Voltage (Reference/ Ranges/ Resolution/ Accuracy/Remarks): 

+ 100 mV/ -10 to 110 mV/ 1 mV/ 0.02 + 0.01/ Maximum Output: 0.5 mA; Maximum Output: 2 mA

+ 1000 mV/ -100 to 1100 mV/ 10 mV/ 0.02 + 0.01/ Maximum Output: 0.5 mA; Maximum Output: 2 mA

- Resistance (Ohm) (Reference/ Ranges/ Resolution/ Accuracy/Remarks): 

+ 400 Ω/ 0 to 400 Ω/ 0.01 Ω/ 0.02 + 0.02/ _

+ 4 KΩ/ 0 KΩ to 4 KΩ/0.1 Ω/ 0.05 + 0.025/ _

- Thermocouple (Reference/ Ranges/ Resolution/ Accuracy/Remarks):

+ R: 0 to 1767°C/ 1°C/ 0 to 100°C: 1.5°C/ ITS-90

                     / 100 to 1767°C: 1.2°C/

+ S: 0 to 1767°C/ 1°C/ 0 to 100°C: 1.5°C/ ITS-90

                     /100 to 1767°C: 1.2°C/

+ K: -200 to 1372°C/ 0.1°C/ -200 to -100°C: 0.6°C/ ITS-90

                          /-100 to 400°C : 0.5°C/

                          /400 to 1200°C: 0.7°C/

                          /1200 to 1372°C: 0.9°C/



+ E: -200 to 1000°C/ 0.1°C/ -200 to 100°C: 0.6°C/ ITS-90

                          / -100 to 600°C: 0.5°C/

                          / 600 to 1000°C: 0.4°C/

+ J: -200 to 1200°C/ 0.1°C/ -200 to -100°C: 0.6°C/ ITS-90

                          / -100 to 800°C: 0.5°C/

                          / 800 to 1200°C: 0.7°C/

+ T: -250 to 400°C/ 0.1°C/ -250 to 400°C: 0.6°C/ ITS-90

+ N: -200 to 1300°C/ 0.1°C/ -200 to -100°C: 1.0°C/ ITS-90

                          / -100 to 900°C: 0.7°C/

                          / 900 to 1300°C: 0.8°C/


+ B: 600 to 1820°C/ 1°C/ 600 to 800°C: 1.5°C; 800 to 1820°C: 1.1°C/ ITS-90

+ L: -200 to 900°C/ 0.1°C/ -200 to 0°C: 0.7°C; 0 to 900°C: 0.5°C/ ITS-90

+ U: -200 to 600°C/ 0.1°C/ -200 to 0°C: 0.7°C; 0 to 600°C: 0.5°C/ ITS-90

 


- RTD (Reference/ Ranges/ Resolution/ Accuracy):

+ Pt100 385/ -200 to 800°C/ 0.1°C/ -200 to 0°C: 0.3°C

                                 / 0 to 400°C: 0.5°C

                                 /400 to 800°C: 0.8°C

+ Pt1000 385/ -200 to 630°C/ 0.1°C/ -200 to 100°C: 0.8°C

                                 / 100 to 300°C: 0.9°C

                                 /300 to 630°C: 1.0°C

+ Pt200 385/ -200 to 630°C/ 0.1°C/ -200 to 100°C: 0.4°C

                                 / 100 to 300°C: 0.5°C

                                 /300 to 630°C: 0.7°C

+ Pt500 385/ -200 to 630°C/ 0.1°C/ -200 to 100°C: 0.2°C

                                 / 100 to 300°C: 0.5°C

                                 /300 to 630°C: 0.7°C


+ Cu10/ -100 to 260°C/ 0.1°C/ 1.8°C

+ Cu50/ -50 to 150°C/ 0.1°C/ 0.6°C

Standard delivery:

Process Meter – Model : Nagman 05+

1.5V ‘AAA’ Alkaline Batteries (4 Nos.)

Operation Manual

Test Leads

Traceable Calibration Certificate

Optionals:

Calibration Certificate from NABL Accredited –as per ISO/IEC 17025:2005 Laboratory 


Details

Datasheet