For full functionality of this site it is necessary to enable JavaScript.

PRESYS T-650P Dry Blocks for Temperature Calibration (from ambient temperature to 650 °C; ± 0.05 °C)

Operating Range (ambient temperature: 23 °C): from ambient temperature to 650 °C

Accuracy 

Internal reference: ± (0.1 °C + 0.1% of reading)

with external probe:± 0.10 °C

Resolution: 0.01 °C

Stability: ± 0.05 °C

Axial Uniformity (40 mm):

± 0.05 °C @ 50 °C

± 0.10 °C @ 300 °C

± 0.20 °C @ 650 °C

Radial Uniformity (homogeneity):

± 0.05 °C @ 50 °C

± 0.10 °C @ 300 °C

± 0.20 °C @ 650 °C

Heating Time:  18 min (50 °C to 650 °C)

Cooling Time: 22 min (650 °C to 100 °C)

Power Supply:115 or 230 Vac, 50/60 Hz

Electric Power: 1000W

Display: Graphic vacuum fluorescent with contrast adjustment

Well Diameter x Depth: Ø 32 mm x 124 mm

Dimension (HxWxD): 260 x 180 x 270 mm

Weight: 10 kg