Measuring range and Accuracy (at 23℃±5℃) *D<0.1 parallel equivalent circuit
Test Current/ Standard accuracy/ Measuring voltage
20pF/ within ±0.25% of rdg±2digit±0.01pF/ 1V
200pF/ within ±0.2% of rdg±2digit/ 1V
2nF/ within ±0.2% of rdg±2digit/ 1V
20nF/ within ±0.2% of rdg±2digit/ 1V
200nF/ within ±0.2% of rdg±2digit/ 1V
2μF/ within ±0.2% of rdg±2digit/ 1V
20μF/ within ±0.2% of rdg±2digit / less than 1V
200μF/ within ±0.8% of rdg±3digit [※within 100μF] / less than 1V
※More than 100μF is (±0.8% of rdg±3digit)×measurement value[μF]/within 100
Measuring method: 5-terminal measurement, parallel equivalent circuit
Measuring frequency: 1kHz ±0.1%, SINE wave
Stray capacity correcting range: about 20pF
Full scale and zero temp. coefficiency: within ±100ppm/℃
Measuring time:
[Remote start mode] 1.2msec.
[Free running] 5~10 times per second
Comparator set range: 1999 both HI and LO
Operation condition: [Temp.] 5*C~+40*C [Humidity] less than 85%
Power supply: AC100V~240V selectable, 50/60Hz, about 30VA
Outer dimension: about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.)
Weight: about 5.8kg