Common problems in old sterilizers that need to be watched out for

06/20/2025 09:08:36

Used autoclaves that have been operated for a long time will often encounter some unavoidable problems. So when encountering common problems with old autoclaves, what should users pay attention to and how should they handle them? EMIN invites you to learn more in the article below.

Common problems in old sterilizers after long-term use

Autoclaves are indispensable equipment in microbiology laboratories, often used to sterilize specimens and instruments at high temperatures (such as 121°C), helping to prevent contamination and ensure accurate experimental results. 

However, after a long period of operation, old autoclaves that have been used for many years often begin to develop errors, causing many inconveniences during operation. So what are these common problems?

Damaged resistor

The resistor is the main heating element in the autoclave, ensuring that the internal temperature always meets the requirements. After a long period of operation, the resistor may be corroded, covered with scale or accumulated with dust, reducing the heat generating capacity of the device.

When the resistor no longer ensures the heating function, the temperature and pressure control in the autoclave chamber may be distorted, affecting the sterilization efficiency as well as the overall operation.

Aging rubber seal

The rubber gasket surrounding the lid of the autoclave acts to prevent heat loss during the sterilization process. In equipment that has been used for many years, the rubber gasket can age and lose its elasticity, leading to heat leakage and reduced operating efficiency. Heat escaping slows down the sterilization process and poses a potential safety risk to the operator.

Lid lock is broken

In addition to the rubber gasket, the lid lock is also the part that keeps the pot lid tightly closed during operation. After many years of use, the lock on old sterilizers can be damaged, making the lid unable to close completely, thereby causing unsafe operation of the device.

Leaks at valves and pipes

Leaks in the valve or pipe of the autoclave will cause the device to not maintain the necessary tightness, directly affecting its ability to operate. Not only will the required pressure and temperature not be achieved, this problem also increases the risk of overheating and causing danger during use.

Related Posts: 

Is there any harm in using an autoclave?

How to use an autoclave and what are the uses you don't know?

Top 3 most popular autoclaves in 2023

Some things to note when operating old, long-used autoclaves

In order for autoclaves that have been used for more than 7 years to maintain good performance and provide stable sterilization results at a stable temperature, regular maintenance and inspection are essential. So what should operators pay attention to when using old equipment? 

- Regularly remove excess water remaining in the steam chamber every day and use purified water to minimize the risk of damage to the resistor. When the resistor is still good, the heat generation capacity will be stable, helping the device maintain the appropriate pressure and temperature throughout the operating cycle.

- Periodically clean the steam chamber and the pot lid every month to avoid dust accumulation, affecting the operation process. When detecting signs of rubber gasket loss or not ensuring tightness, it is necessary to replace it immediately to avoid heat loss and ensure the device always operates effectively. 

- Regularly check the tightness of the locking pins on the lid to ensure that the pot is always closed during operation. In addition, the lid closing operation must also be performed in the correct order to avoid breaking or warping the locking mechanisms.

- It is necessary to periodically monitor the temperature and pressure parameters at various points inside the autoclave using a specialized data logger with high accuracy.

- For autoclaves with integrated vacuum function, it is recommended to conduct leak testing, temperature and pressure control according to international standards such as UNI EN ISO 13060, using specialized equipment such as Tecnosoft PressureDisk Datalogger for accurate results.

Related News

Which Solar PV Safety Standards Apply in Vietnam?
09/28/2026 09:41:38

A solar PV system may use certified solar panels, inverters, and protective devices, but this does not automatically mean that the entire system is safe. The installation quality of the complete system must be independently checked against the applicable safety requirements.

What Are Push-to-Start Drivers Used For in Assembly?
09/26/2026 11:19:18

Push-to-start drivers start when the operator applies deliberate pressure to the workpiece. They can support more natural hand movement, repeatable torque application, reduced accidental starts, and faster fastening cycles in assembly, manufacturing, and maintenance work.

How Do Preset Torque Wrenches Improve Assembly Lines?
09/26/2026 11:17:32

Preset torque wrenches support consistent fastening in repetitive assembly by using a fixed torque setting. Selection should consider fastener size, required accuracy, and special assembly requirements.

How Much Power Can a Rooftop Solar System Still Generate After 5 Years?
09/26/2026 08:44:51

Sau 5 năm vận hành, hệ thống điện mặt trời áp mái có thể vẫn hoạt động ổn định nhưng sản lượng không còn giống thời điểm mới lắp đặt. Mức suy giảm phụ thuộc vào tuổi thọ tấm pin, nhiệt độ, bức xạ, độ ẩm, bụi bẩn, dây dẫn, đầu nối và bộ biến tần.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts