Digital MultiMeter Calibration Procedure

04/06/2021 10:26:04

**1.Range of application** \- DC Voltage 1000V \- AC Voltage 1000V, Frequency 100kHz \- DC Current 30A \- AC Current 30A, Frequency 10kHz \- Resistor 100

1.Range of application

- DC Voltage 1000V

- AC Voltage 1000V, Frequency 100kHz

- DC Current 30A

- AC Current 30A, Frequency 10kHz

- Resistor 1000MΩ

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s): Multifunction Calibrator (MFC)

Meatest M142 Multifunction Calibrator 15 ppm

- DC Voltage: 0µV - 1000V.

- AC Voltage: 0,1mV - 1000V, Frequency: 20Hz - 100kHz.

- DC Current: 0µA - 30A.

- AC Current: 1µA - 30A, Frequency: 20Hz - 10kHz.

- Resistor: 0 Ω - 1GΩ.

3.2 Auxiliary Instrument(s)

3.2.1 Testo Hygro-Thermometer 608-H1

Testo 608-H1 - Thermal Hygrometer (Part Number 0560 6081) | TEquipment

- Range of Measurement: (0 ÷ 50) °C; Accuracy: ±0,5 °C

- Range of Measurement: (10 ÷ 95) %RH; Accuracy: ±3 %RH

3.2.2 Dedicated Measuring Probe

- Probes with small electromotive force.

4. Conditions for Calibration

- Temperature: (23 ± 5) °C.

- Humidity: no higher than 80% RH.

5. Prepare for Calibration

- Make sure the power supply to the DMM meets the requirements specified in the catalog, the fuses, the protective circuit are still working properly.

- The DMM should be placed in the calibration environment for at least 8 hours.

- MFC and DMM must be turned on and dried for at least 30 minutes prior to calibration.

6. Begin the calibration

a. Outside Check

b. Technical Check

c. Measurement Check

    1. General requirements

    2. Begin the DMM Calibration

        a. DC Voltage

        b. AC Voltage

        c. DC Current

        d. AC Current

        e. Resistor

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated DMM is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of DMM is 12 months.


For calibration of DMM, click here.

For further information about Digital MCalibration service, please contact

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220


Related News

EMIN is an official Acepom distributor in Vietnam
11/27/2026 17:18:59

Since January 2021, EMIN Vietnam has served as the official distributor of Acepom products and services in Vietnam. Acepom’s industrial maintenance and diagnostic equipment is specifically engineered and developed to optimize bearing lifespan and enhance the stability of rotating machinery systems across a wide range of industries.

EMIN is an official distributor for 3nh and Tilo in Vietnam
11/27/2026 16:15:47

Since October 2021, EMIN has served as the official distributor for products and services from 3nh and Tilo in Vietnam. These brands are renowned for their color-measurement and optical instruments, which meet international metrological standards for use in both manufacturing and laboratory environments.

Triển lãm "Dây & Cáp điện 2026" - Ngày 2
11/25/2026 14:31:19

Tại gian hàng D02, EMIN mang 4 thiết bị Rohde & Schwarz đến để trưng bày và vận hành trực tiếp: máy phân tích tín hiệu và phổ R&S®FSV3044, máy hiện sóng hiệu năng cao R&S®RTP134, máy phân tích mạng vector R&S®ZNLE4 và máy phân tích mạng vector cầm tay R&S®ZNH18

What Is 5 µs End-to-End Sync Verification in Substations?
11/25/2026 10:55:08

In a 50 Hz digital substation using 80 Sampled Values per cycle, samples are 250 µs apart. End-to-end verification examines whether timing alignment is maintained so protection functions can use comparable measurements.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts