FLUKE Fluke 5624-20-B High-Temperature Probe (0 °C ~ 1000 °C)
Range: 0 °C to 1000 °C
Calibration Uncertainty:
±0.004 °C at 0 °C
±0.020 °C at 962 °C
Long-Term Drift (Rtpw), k = 2:
<0.01 °C at 0 °C / 100 hours at 1000 °C
<0.06 °C at 0 °C / 1000 hours at 1000 °C
Calibrated Accuracy† (k = 2):
±0.006 °C at 0 °C
±0.031 °C at 660 °C
±0.041 °C at 961 °C
Short-Term Stability: ±0.005 °C
Immersion Depth: <153 mm (6 in) at 700 °C
Resistance at the Triple Point of Water (Rtpw): 10 Ω (±0.1 Ω)
Hysteresis: <0.005 °C from 0 °C to 1000 °C
Thermal Cycling: <0.01 °C, 10 cycles from 0 °C to 1000 °C
Current: 1 mA
Size: 6.35 mm (0.25 in) O.D.
Length: 508 mm (20 in)
Sheath Material: Alumina