GEO Calibration Multi-Gen3 Humidity Generator (80% RH, 18 °C ~ 28 °C)

GENERAL SPECIFICATIONS (For Each Chamber - Chamber 1,2 and 3)

OPERATING AMBIENT CONDITIONS

TEMPERATURE : 18 °C to 28 °C

HUMIDITY:  Up to 80 % RH

STORAGE CONDITIONS

TEMPERATURE : -20 °C to 50 °C

HUMIDITY:  0 % to 95 % RH (non-condensing)

ALTITUDE: 2000 m

CALIBRATION SYSTEM ACCURACY

TEMPERATURE : ± 0.12 °C Or Better Typically ± 0.10 °C Based on Probe Accuracy

HUMIDITY : ± 1.00 % RH Or Better Typlically ± 0.80 % Based on Probe Accuracy

CONTROLLER TYPE : PID Controller

UNIFORMITY

TEMPERATURE : 0.10 °C Or Better Typically ± 0.10 °C

HUMIDITY : 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %

STABILITY

TEMPERATURE:  0.05 °C

HUMIDITY:  0.15 % RH @ 18 °C to 28 °C

WARM-UP PERIOD

AMBIENT CONDITIONS : 30 Minutes Maximum

COLD : 30 Minutes

RESOLUTION

DISPLAY

Temperature: 0.01 °C

Humidity : 0.01 % RH

Dew Point : 0.01 °C (Calculated)

USB

Temperature:  0.01 °C

Humidity:  0.01 % RH

RAMP / SOAK RATE OF CHANGE

Depends on ambient and desiccant conditions

TEMPERATURE

 High → Low 1.50 °C / minute *Typical

 Low → High 3.00 °C / minute *Typical

HUMIDITY 

High → Low 5.00 % RH / minute *Typical 

Low → High 10.00 % RH / minute *Typical

CONTROL RANGE
TEMPERATURE 5 °C to 55 °C
HUMIDITY@ 18°C 5 % to 95 % RH HUMIDITY@ 35°C 5 % to 85 % RH
HUMIDITY@ 23°C 5 % to 95 % RH HUMIDITY@ 40°C 5 % to 75 % RH
HUMIDITY@ 30°C 5 % to 90 % RH HUMIDITY@ 50°C 5 % to 50 % RH