REX M600L-2025 Multi-parameter Analyzer (2000.00 mV, 20.000 pH, 20.000 pX, 135.0 ºC, 100.0 MΩ·cm)

ModelM600L-2025
P/N651223CB2
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Specifications


Module: pH module

mV

Range: (-2000.00~2000.00) mV

Resolution: 1 mV, 0.1 mV, 0.01 mV

Accuracy: ±0.03% or ±0.1 mV

pH

Range: (-2.000~20.000) pH

Resolution: 0.1 pH, 0.01 pH, 0.001 pH

Accuracy: ±0.002 pH

Temperature

Range: (-10.0~135.0) ºC, (14.0~275.0) ºF

Resolution: 0.1 ºC, 0.1 ºF

Accuracy: ±0.3 ºC (0.0~60.0) ºC; ±1.0 ºC (Other Range)


Module: pX module

mV

Range: (-2000.00~2000.00) mV

Resolution: 1 mV, 0.1 mV, 0.01 mV

Accuracy: ±0.03% or ±0.1 mV

pH

Range: (-2.000~20.000) pH

Resolution: 0.001 pH

Accuracy: ±0.002 pH

pX

Range: (-2.000~20.000) pX

Resolution: 0.001 pX

Accuracy: ±0.002 pX

Ion

Range: (1.000e-9 ~ 9.999e+9), Unit: mol/L, mmol/L, g/L, mg/L, μg/L, ppm, ppb

Resolution: Up to 4 significant digits

Accuracy: ±0.3%

Temperature

Range: (-10.0~135.0) ºC, (14.0~275.0) ºF

Resolution: 0.1 ºC, 0.1 ºF

Accuracy: ±0.3 ºC (0.0~60.0) ºC; ±1.0 ºC (Other Range)


Module: EC-T module

Conductivity

Range: 0.000 μS/cm ~ 3000 mS/cm

Resolution: 0.001 μS/cm minimum, changed with range

Accuracy: ±0.5% FS

Repeatability: 0.40% FS

Resistivity

Range: 5.00 Ω·cm ~ 100.0 MΩ·cm

Resolution: 0.01 Ω·cm minimum, changed with range

Accuracy: ±0.5% FS

TDS

Range: 0.000 ppm ~ 1000 ppt

Resolution: 0.001 ppm minimum, changed with range

Accuracy: ±0.5% FS

Salinity

Range: (0.0~80.0) ppt, (0.00~8.00) %NaCl, (2.000~42.000) psu

Resolution: 0.1 ppt, 0.1% NaCl, 0.001 psu

Accuracy: ±1 ppt

Temperature

Range: (-10.0~135.0) ºC, (14.0~275.0) ºF

Resolution: 0.1 ºC, 0.1 ºF

Accuracy: ±0.3 ºC (0.0~60.0) ºC; ±1.0 ºC (Other Range)


Module: EC-F module

Conductivity

Range: 50.0 μS/cm ~ 3000 mS/cm

Resolution: 0.001 μS/cm minimum, changed with range

Accuracy: ±1.5% FS

Resistivity

Range: 5.00 Ω·cm ~ 20.0 kΩ·cm

Resolution: 0.01 Ω·cm minimum, changed with range

Accuracy: ±1.5% FS

TDS

Range: 25.0 ppm ~ 1000 ppt

Resolution: 0.001 ppm minimum, changed with range

Accuracy: ±1.5% FS

Salinity

Range: (0.0~80.0) ppt, (0.00~8.00) %NaCl, (2.000~42.000) psu

Resolution: 0.1 ppt, 0.1% NaCl, 0.001 psu

Accuracy: ±2 ppt

Temperature

Range: (-10.0~135.0) ºC, (14.0~275.0) ºF

Resolution: 0.1 ºC, 0.1 ºF

Accuracy: ±0.5 ºC (0.0 ºC~60.0 ºC); ±1.5 ºC (Other Range)


Module: DO-P module

Dissolved Oxygen Concentration

Range: (0.00~99.99) ppm

Resolution: 0.01 ppm

Accuracy:

±0.10 ppm, (0.00~50.00) ppm

±0.50 ppm, (50.0~99.99) ppm

Repeatability: 0.15 ppm

Zero Accuracy: ≤0.1 ppm

Instrument Accuracy:

≤20.00 ppm: ±0.30 ppm

>20.00 ppm: ±10.0%

Response time: ≤45s (20 ºC up to 90% response)

Salinity Accuracy: ±2%

Saturation

Range: (0.0~600.0)%

Resolution: 0.10%

Accuracy: ±2.0%

Temperature

Range: (-10.0~135.0) ºC, (14.0~275.0) ºF

Resolution: 0.1 ºC, 0.1 ºF

Accuracy: ±0.3 ºC (0.0~60.0) ºC; ±1.0 ºC (Other Range)


Module: DO-F module

Dissolved Oxygen Concentration

Range: (0.00~20.00) ppm

Resolution: 0.01 ppm

Repeatability: 0.15 ppm

Zero Accuracy: ≤0.1 ppm

Instrument Accuracy: ±0.30 ppm

Response time: ≤60s (20 ºC, up to 90% response)

Salinity Accuracy: ±2%

Saturation

Range: (0.0~200.0)%

Resolution: 0.10%

Accuracy: ±10.0%

Temperature

Range: (0.0~50.0) ºC, (32.0~122.0) ºF

Resolution: ºC, ºF

Accuracy: ±0.3 ºC (0.0~50.0) ºC