VELP TB1 Turbidimeter (0 ~ 1000 NTU)

Construction material: ABS

Measurement method: ISO 7027 compliant nephelometric method (90°)

Calibration standard: 0.02, 20, 100, 800 NTU

Light source: Infrared emitting-diode (850 nm wavelength)

Protection degree CEI EN 60529: IP 67

Weight: 0.2 Kg (0.4 lb)

Dimensions (WxHxD): 68x50x155 mm (2.7x2x6.1 in)

Measurement range: from 0 to 1000 NTU

Accuracy:

±2% in the range from 0 to 500 NTU

±3% in the range from 501 to 1000 NTU

Repeatability:±0.01 NTU or ±1% of reading, whichever is greater with gel samples