Direct access to our certified experts
Channels:2 CH
Pulse Period:
• Range: 5ns to 8 sec.
• Resolution: 10 ps
• Jitter: 15 ps
Max Pulse Frequency: 400 MHz (Double, Triple, and Quad Mode)
Accuracy: ±2 ppm
Pulse Mode: Single, Double, Triple, Quadruple Pulse; External Width
Pulse Width:
• Range: 1 ns to 8 sec.
• Resolution: 10 ps;
• Accuracy: ± (1 % + 100 ps)2
• Jitter: < 10ps
Pulse Delay:
• Range: 0 ps to 8 secs
• Resolution: 10 ps
• Accuracy: ± 0.1% + 100ps
Pulse Amplitude:
• Range pk-pk: 100 mVpp to ±50 Vpp
• Resolution: 10 mV
Baseline Offset:
• Range: -25V, -12,5V, 0V, +12,5V, +25 V (software selectable)
• Resolution: 12.5V
Rise and Fall Times:
• Rise/Fall Time (20% to 80%, Offset=0V): < 380 ps (up to 25Vpp amplitude)
• Rise Time (20% to 80%, Offset=0V): < 400 ps; Fall Time (20% to 80%, Offset=0V): < 490 ps
• Rise/Fall Time (10% to 90%, Offset=0V): < 560 ps (up to 25Vpp amplitude)
• Rise Time (10% to 90%, Offset=0V): < 600 ps; Fall Time (10% to 90%, Offset=0V): < 800 ps
Overshoot: < 8 % (up to 25Vpp amplitude); < 5 %
Channel to Channel Jitter: < 15 ps
Output Impedance: 50 Ω
Trigger Mode: Single, Continuous, Burst, Gated
Trigger Input Impedance: 50 Ω or 1kΩ User Selectable
Trigger Input Range: ± 3.5 V (50 Ω Input Impedance); ± 10 V (1KΩ Input Impedance); < 50mVpp Minimum Detectable Amplitude
Trigger Threshold: ± 8V; 10 mV Resolution; ± 100 mV Accuracy
Maximum Trigger Frequency: 40 MHz
Minimum Trigger Pulse Width: 1ns
Trigger Jitter: < 30ps (Trig IN to Trig Out) with Frequency ≤ 15 MHz
Reference Clock Input:
• 10MHz to 100MHz;
• 50 Ω Input Impedance (AC Coupled);
• -5 dBm to 4 dBm Sine or Square wave ( RT <1 ns and Duty Cycle between 40% to 60%)
Reference Clock Output:
• 10MHz or Reference Clock Input Frequency;
• 50 Ω Output Impedance (DC Coupled);
• Adjustable Amplitude Square wave: 0V to 1.25 V into 50 Ω [ 0V to 2.5 V into Hi-Z ]
Power Input: 100 – 240 VAC ± 10%, 47 – 63 Hz,
Power Consumption: 300 Watts
Environment:
• Operating: 41°F to +104 °F [ 5 °C to +40 °C ]; 5% to 80 % RH Non-condensing; 9,842 feet [ 3,000 meters]
• Storage: -4 °F to 140 °F [-20 °C to +60 °C ]; 5% to 95 % RH Non-condensing; 39,370 feet [ 12,000 meter]
EMC and Safety: EN61010-1 (Safety); EN 61326-1:2013 (Standards); EN 61326-1:2013 (Immunity)
Hardware and Software: Intel Celeron 2GHz or Better; 8 GB RAM; 120 GB SSD; Windows 10; 7″ Touch LCD; HDMI, VGA, USB, RS-232, and LAN Ports
Dimension (W x H x D): 17.5 x 5.3 x 12.6 ” [ 445 x 135 x 320 mm ] (3U 19” rackmount)
Weight: 21.7 lbs [ 9.7 kg ]