Power rating: 0 – 500 W (RMS)
Accuracy: ±(1 W + 5% of reading)
Duty cycle: 100% up to 60 seconds
Load bank: 0 – 5115 Ω
Resolution: 5 Ω
Voltage (peak): 0 – 10 kV (Peak to Peak) – Closed load only
Accuracy: ±(10% of reading + 15 V)
Note: Measurement is taken between active and dispersive electrodes with closed load only
Voltage: 0 – 700 V (RMS)
Accuracy: ±(10% of reading + 5 V)
Current (RMS): 0 – 6000 mA with load bank; 0 – 8000 mA external load test
Accuracy: ±(2% of reading + 10 mA)
Crest factor: 1.4 – 20 (Vpeak / VRMS)
Calculation: The higher of the two peak voltage measurements is used
RMS Bandwidth: 30 Hz to 10 MHz (-3 dB)
With loads: 30 Hz to 2.5 MHz (-3 dB)
Variable loads: 5 – 5115 Ω, steps @ 5 Ω (1023 steps)
Accuracy: ±(1%, +0.5, -0.0 Ω of set load)
Load array: Ceramic resistors (non-inductive)
Measurement Delay
Measurement delay: Foot switch delay selectable 200 – 5000 ms (10 ms resolution)
RF Leakage (High Frequency Leakage)
Active: From active part to earth
Passive: From plate receptacle to earth
Load: Variable (see power measurement) / Fixed 2 x 200 Ω
Accuracy: ±1%, +0.5, -0.0 Ω
Contact Quality Monitoring (CQM)
Range: 0 – 475 Ω, steps @ 1 Ω
Mechanism: Motor driven potentiometer
Accuracy: ±5% ± 2 Ω
Alarm register: High and low, manual confirmation
Ranging: Manual or automatic
Output Connectors
Remote foot switch control (CUT): 2 x 4 mm – yellow, single relay contact
Remote foot switch control (COAG): 2 x 4 mm – blue, single relay contact
High frequency leakage: Through 4 mm sockets and power measurements
USB: PC download
Oscilloscope output: 0.5 V/A, minimum 100 mA RF current (indication only, un-calibrated)
Isolation: 10 kV isolation between measurement device and enclosure
Low frequency filter: 100 Hz filter
General
Memory: Approx. 5,000 records (4 MB)
Output: CSV and SSS format
Dimensions: 370 × 300 × 204 mm (14.6” × 11.8” × 8”)
Weight: 10 kg (22 lbs)
Operating temperature: 10°C – 40°C
Storage temperature: 0°C – 50°C
Mains power: 115 / 230 V AC ±10%, 48 – 66 Hz, 35 VA
Fuses: 2 × 1.6 A (T) ceramic