Scale Calibration Procedure

04/26/2021 13:44:25

**1\. Range of application** This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not gre

1. Range of application

This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not greater than 1 mg. Technical scale is (non-automatic scale class II) according to OIML R76.

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s)

- Analytical Balance:
Kern ADB 200-4 Analytical balance (200g/0.1mg)
Sets of weights E2, Class E2, Measuring range: 1 mg ÷ 500g  
Sets of weights F1, Class F1, Measuring range: 1g ÷ 500g
- Technical Scale:

Sets of weights F1, Class F1, Measuring range 1g ÷ 500g
Standard Weight F1, Class F1, Measurement weight: 1 kg, 2 kg, 5 kg, 10 kg.
- Bench Scale:

Standard Weight M11, Class M1, Measuring range 0 ÷ 150kg

3.2 Auxiliary Instrument(s)
Hygro - Thermometer:

- Range of Measurement: (0 ÷ 50) °C
- Accuracy: ± 0,5 °C
- Range of Measurement: (10 ÷ 95) %RH
- Accuracy: ± 3 %RH

4. Conditions for Calibration
The calibration site must be bright, away from heat sources, wind sources, and not vibrated.
Temperature, humidity in the calibration site ensures:
Ambient temperature: (22 ± 3) ° C
Air humidity: (40 ÷ 80) %RH.
The scale should be calibrated at the site of use.

5. Prepare for Calibration
Before calibration, the following preparations must be done:
- The scale must be clean.
- Place the scale to be calibrated firmly on a surface and adjust the scale.
- Turn on the scale for at least 30 minutes or as required by the manufacturer.
- The scale to be calibrated should be thermally equilibrium with the environment in which it is calibrated. The minimum time required for this process is specified by the manufacturer.

6. Begin the calibration
6.1 Outside check
6.2 Technical Check
6.3 Measurement check
6.3.1 General requirements
6.3.2 Begin the calibration
- Test repeatability
- Check the deviation angle
- Check the correctness

7. Measurement Uncertainty Evaluation
8. General handling
- The calibrated scale is stamped, certified in paper.
- The calibration result includes standard value, indicator value, main number, together with the measurement uncertainty expanded for each weight level, coverage factor k.
- The recommended calibration cycle of scale is 12 months.

9. About EMIN's Calibration - Verification - Repair center
EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)
If you want your instruments calibrated, please contact:
Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 
Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City
Email: services@emin.vn - Hotline: 0968.060.220
Further information about scale calibration service can be found here.


Related News

Which Vacuum Pump Should You Choose for a Microbiology Lab to Avoid Affecting Samples?
07/15/2026 14:45:07

Vacuum pumps are utilized in a wide range of microbiology laboratory procedures, from membrane filtration and the aspiration of supernatants following centrifugation to sample concentration and the support of negative-pressure systems. Selecting the appropriate pump not only ensures more stable operational processes but also minimizes the risk of sample contamination, reduces maintenance costs, and extends the equipment's lifespan

QCVN 25:2025/BKHCN Takes Effect: Do Businesses Need to Prepare Electrical Safety Testing Equipment?
07/15/2026 09:15:04

Electrical safety is consistently one of the primary criteria evaluated before electrical or electronic equipment—or industrial components—are released to the market. A product may operate reliably yet still fail to meet requirements if its insulation lacks sufficient durability, its grounding resistance exceeds permissible limits, or leakage current occurs during operation.

Why Is Associated Research Considered an Icon of Hipot Technology?
07/14/2026 09:05:30

When it comes to electrical safety testing equipment specifically Hipot testers—many QA/QC and testing engineers are familiar with the globally renowned brand Associated Research. A subsidiary of the parent company IKONIX, Associated Research (ARI) invented the world's first commercial Hipot tester.

What Is the Standard 3-Phase Voltage? How to Read Cabinet-Mounted Meters Without Errors
07/13/2026 15:09:18

Voltmeters are almost invariably found on the panels of electrical cabinets in factories, power stations, or generator systems. A glance lasting just a few seconds allows an operator to determine whether the power supply is stable or showing signs of abnormality. However, not everyone understands the significance of the displayed readings, particularly regarding three-phase electrical systems.

Distinguishing between ARI’s Hypot, HypotUltra, and HypotMax: Which device is right for your factory?
07/13/2026 09:48:30

Few people realize that the concept of the "Hipot Test"—or electrical safety testing—has long existed within the electrical equipment manufacturing industry. Driven by the growing need to verify the voltage withstand capabilities of conductors, motors, and electrical devices, the industry required a robust and virtually fail-safe method to detect insulation weaknesses before products reached the market.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts